§15729. Laboratory Quality Assurance (LQA) Program.
(a) The licensed laboratory shall develop and implement a LQA program to assure the reliability and validity of the analytical data produced by the laboratory. The LQA program shall, at minimum, include a written LQA manual that addresses the following:
- Quality control procedures;
- Laboratory organization and employee training and responsibilities, including good laboratory practice (GLP);
- LQA objectives for measurement data;
- Traceability of data and analytical results;
- Instrument maintenance, calibration procedures, and frequency;
- Performance and system audits;
- Corrective action procedures;
- Steps to change processes when necessary;
- Record retention and document control;
- Test procedure standardization; and
- Method validation.
(b) The supervisory or management laboratory employee shall annually review, amend if necessary, and approve the LQA program and manual both when they are created and when there is a change in methods, laboratory equipment, or the supervisory or management laboratory employee.
Authority cited: Section 26013, Business and Professions Code.
Reference: Sections 26100, 26104 and 26110, Business and Professions Code.
§15730. Laboratory Quality Control (LQC) Samples.
The licensed laboratory shall use LQC samples and adhere to good laboratory practice (GLP) in the performance of each analysis according to the following specifications.
(a) The licensed laboratory shall analyze LQC samples in the same manner as the laboratory analyzes cannabis and cannabis products samples.
(b) The licensed laboratory shall use at least one negative control, one positive control, and one laboratory replicate sample in each analytical batch for each target organism during microbial testing. If one of the controls produces unexpected results, the samples shall be re-prepped and reanalyzed with a new set of controls.
(c) If the result of the microbial analyses is outside the specified acceptance criteria in the following table, the licensed laboratory shall determine the cause and take steps to remedy the problem until the result is within the specified acceptance criteria.
Laboratory Quality Control Sample | Acceptance Criteria | Corrective Action |
Positive control | Produces expected result, positive result | Re-prep and reanalyze the entire analytical batch, once. If problem persists, locate and remedy the source of unexpected result, then re-prep samples and reanalyze with a new set of controls. |
Negative control | Produces expected result, negative result | Re-prep and reanalyze the entire analytical batch, once. If problem persists, locate and remedy the source of unexpected result, then re-prep samples and reanalyze with a new set of controls. |
Laboratory replicate sample | Sample results must concur | Reanalyze sample and associated replicate sample once. If problem persists, re-prep samples and reanalyze. |
(d) The licensed laboratory shall prepare and analyze at least one of each of the following LQC samples for each analytical batch:
- Method blank;
- Laboratory control sample (LCS); and
- Laboratory replicate sample or matrix spike sample.
(e) The laboratory shall analyze, at minimum, a continuing calibration verification (CCV) sample at the beginning of each analytical sequence and every 10 samples thereafter.
(f) If the result of the chemical analyses is outside the specified acceptance criteria in the following table, the laboratory shall determine the cause and take steps to remedy the problem until the result is within the specified acceptance criteria.
Laboratory Quality Control Sample | Acceptance Criteria | Corrective Action |
Method Blank sample | Not to exceed LOQ | Reanalyze entire analytical batch once. If method blank is still greater than the LOQ for any analyte, locate the source of contamination then re-prep samples and reanalyze. |
LCS | Percent recovery 70%to 130% | Reanalyze the entire analytical batch, once. If problem persists, re-prep samples and reanalyze or re-run the initial calibration curve. |
Laboratory replicate sample | RPD ≤30% | Reanalyze sample and associated replicate sample once. If problem persists, re-prep samples and reanalyze. |
Matrix spike sample | Percent recovery between 70% to 130% | Reanalyze sample and associated matrix spike sample once. If problem persists, re-prep samples and reanalyze. |
CCV | Percent recovery between 70% to 130% | Reanalyze all samples that followed the last CCV that met the acceptance criteria. If CCV still fails, re-run the initial calibration curve and all samples in the analytical sequence. |
(g) If any analyte is detected above any action level, as described in this chapter, the sample shall be re-prepped and reanalyzed in replicate within another analytical batch.
- For quantitative analyses, the re-prepped sample and its associated replicate must meet the acceptance criteria of RPD ≤30%.
- For qualitative analyses, the re-prepped sample and its associated replicate results must concur.
(h) If any LQC sample produces a result outside of the acceptance criteria, the laboratory cannot report the result and the entire batch cannot be released for retail sale. The laboratory shall determine the cause and take steps to remedy the problem until the result is within the specified acceptance criteria.
(i) If the licensed laboratory determines that the result is a false-positive or a false- negative, the Department may ask for the laboratory to re-sample or retest.
(j) The licensed laboratory shall compile and generate one LQC sample report for each analytical batch that includes LQC acceptance criteria, measurements, analysis date, and matrix.
Authority cited: Section 26013, Business and Professions Code.
Reference: Sections 26100, 26104 and 26110, Business and Professions Code.
§15731. Limits of Detection (LOD) and Limits of Quantitation (LOQ) for Quantitative Analyses.
(a) The licensed laboratory shall calculate the LOD for chemical method analyses according to any of the following methods:
- Signal-to-noise ratio of between 3:1 and 2:1;
- Standard deviation of the response and the slope of calibration curve using a minimum of 7 spiked blank samples calculated as follows; LOD = (3.3 x standard deviation of the response) / slope of the calibration curve; or
- A method published by the United States Food and Drug Administration (USFDA) or the United States Environmental Protection Agency (USEPA).
(b) The licensed laboratory shall calculate the LOQ for chemical method analyses according to any of the following methods:
- Signal-to-noise ratio of 10:1, at minimum;
- Standard deviation of the response and the slope using a minimum of 7 spiked blank samples calculated as follows:
LOQ = (10 x standard deviation of the response) / slope of the calibration curve; or
- A method published by the USFDA or the USEPA.
Authority cited: Section 26013, Business and Professions Code.
Reference: Sections 26100, 26104 and 26110, Business and Professions Code.
§15732. Data Package.
(a) The licensed laboratory shall compile and generate one data package for each representative sample that the laboratory analyzes.
(b) The licensed laboratory shall create a data package and use the Data Package Cover Page and Checklist Form, DCC-LIC-024 (Amended 2/22), which is incorporated herein by reference. The data package and form DCC-LIC-024 (Amended 2/22) shall be provided to the Department immediately upon request.
Authority: Section 26013, Business and Professions Code.
Reference: Sections 26100, 26104, 26110 and 26160, Business and Professions Code.
§15733. Required Proficiency Testing.
(a) The licensed laboratory shall participate in a proficiency testing program provided by an organization that operates in conformance with the requirements of ISO/IEC 17043, at least once every six months.
(b) The licensed laboratory shall annually, successfully participate in a proficiency testing program for each of the following test methods:
- Cannabinoids;
- Heavy metals;
- Microbial impurities;
- Mycotoxins;
- Residual pesticides;
- Residual solvents and processing chemicals; and
- If tested, terpenoids.
(c) The licensed laboratory shall report all analytes available by the proficiency testing program provider and for which the licensee is required to test as required under this chapter.
(d) The licensed laboratory shall participate in the proficiency testing program by following the laboratory’s existing SOPs for testing cannabis and cannabis products.
(e) The licensed laboratory shall rotate the proficiency testing program among the laboratory employees who perform the test methods.
(f) Laboratory employees who participate in a proficiency testing program shall sign the corresponding analytical reports or attestation statements to certify that the proficiency testing program was conducted in the same manner as the laboratory tests of cannabis and cannabis products.
(g) A supervisory or management laboratory employee shall review and verify the accuracy of results reported for all proficiency testing program samples analyzed.
(h) The licensed laboratory shall request the proficiency testing program provider to send results concurrently to the Department, if available, or the laboratory shall provide the proficiency testing program results to the Department within 3 business days after the laboratory receives notification of their test results from the proficiency testing program provider. Any results shall be reported by submitting the Licensee Notification and Request Form, Notifications and Requests Regarding Testing Laboratories, DCC-LIC- 029 (New 2/22), which is incorporated herein by reference.
Authority: Section 26013, Business and Professions Code.
Reference: Sections 26100 and 26110, Business and Professions Code.
§15734. Satisfactory and Unsatisfactory Proficiency Test Performance.
(a) The licensed laboratory shall be deemed to have successfully participated in a proficiency testing program for an analyte tested in a specific method if the test results demonstrate a “satisfactory” or otherwise proficient performance determination by the proficiency testing program provider.
(b) The licensed laboratory may not report test results for analytes that are deemed by the proficiency testing program provider as “unacceptable,” “questionable,” “unsatisfactory”, or otherwise deficient.
(c) The licensed laboratory may resume reporting test results for analytes that were deemed “unacceptable,” “questionable,” “unsatisfactory”, or otherwise deficient, only if both of the following conditions are met:
- The licensed laboratory satisfactorily remedies the cause of the failure for each analyte; and
- The licensed laboratory submits, to the Department, a written corrective action report demonstrating how the laboratory has fixed the cause of the failure.
Authority cited: Section 26013, Business and Professions Code.
Reference: Sections 26100 and 26110, Business and Professions Code.
§15735. Laboratory Audits.
(a) The licensed laboratory shall conduct an internal audit at least once per year or in accordance with the ISO/IEC 17025 accrediting body’s requirement, whichever is more frequent.
(b) The internal audit must include all of the components required by the ISO/IEC 17025 internal-audit standards.
(c) Within 3 business days of completing the internal audit, the licensed laboratory shall submit the results of the internal audit to the Department.
(d) Within 3 business days of receiving the accrediting body onsite audit findings, the licensed laboratory shall submit the results to the Department.
(e) The licensed laboratory shall submit any audit results to the Department, accompanied by the Licensee Notification and Request Form, Notifications and Requests Regarding Testing Laboratories, DCC-LIC-029 (New 2/22), which is incorporated herein by reference.
Authority: Section 26013, Business and Professions Code.
Reference: Sections 26100 and 26104, Business and Professions Code.